DETERMINATION OF THE DIFFERENCE IN LAG TIME BETWEEN AN ELECTRONIC TEMPERATURE MEASUREMENT SYSTEM AND A MERCURY-IN-GLASS THERMOMETER

ChemicalFormulas - The response time of an electronic temperature measuring device is inherently more rapid than that of a mercury-in-glass thermometer. The temperature measuring device assembly in general use, consisting of the sensor and its casing, or an electronic system and its associated software, or both, is so designed that the temperature measuring system will simulate the temperature lag of the mercury-in-glass thermometer.



To determine the difference in lag time between such a temperature measuring system and a mercury-in-glass thermometer, analyze a sample such as gasoline, kerosine, jet fuel, or light diesel fuel with the electronic temperature measurement system in place and in accordance with the procedures described in this test method. In most cases this is the standard distillation step performed with an automated unit.

  • Do not use a single pure compound, a very narrow boiling range product, or a synthetic blend of less than six compounds for this test.
  • Best results are obtained with a sample that is typical of the sample load of the laboratory. Alternatively, use a full-range mixture with a 5 to 95% boiling range of at least 100°C.

Replace the electronic temperature measuring device with a low range or a high range mercury-in-glass thermometer, depending on the boiling range of the sample.

fluke 1551 ex stik electronic themperature measurement

Repeat the distillation with this thermometer, and manually record the temperature at the various percent recovered 

Calculate the values for the repeatability for the observed slope (DT/DV) for the different readings in the test.

Compare the test data obtained using these two temperature measuring devices. The difference at any point shall be equal to, or less than, the repeatability of the method at that point. If this difference is larger, replace the electronic temperature measuring device or adjust the electronics involved, or both.


Tag : digitale thermometer, infrarood thermometer, thermometer kopen, thermometer buiten, koelkast thermometer, buiten thermometer, binnen buiten thermometer, soorten thermometers, oor thermometer, thermometer koorts, thermometer oor, medisana infrarood thermometer, draadloze thermometer, digitale temperatuurmeter, oorthermometer, hygrometer, vlees thermometer, digitale thermometer baby, hygrometer kopen, koortsthermometer, geijkte thermometer, thermometer digitaal, analoge thermometer, bbq thermometer, thermometers, vleesthermometer, oorthermometer test, temperatuurmeter, barometer, buitenthermometer, termometer, thermometer digital, thermometer infrared, hygrometer digital, bbq thermometer, thermometer baby, braun thermometer, water thermometer, ir thermometer, min max thermometer, baby thermometer, weber thermometer, laser thermometer, oven thermometer, thermometer water, barbecue thermometer, contact thermometer, aquarium thermometer, thermometer hygrometer, thermometer bbq, cdn thermometer, auto thermometer, omron thermometer, thermometer braun, usb thermometer, digital thermometer, thermometer min max, infrared thermometer, thermometer app, minimum maximum thermometer, haccp thermometer, wifi thermometer,